Professor Paul Morgan has developed the following slides demonstrating how to find the default susceptibility-weighted imaging (SWI) sequences on each company's MR scanners:
Recommendations on MRI protocol for CVS and PRL detection
Supplementary Table 2, from 2024 MAGNIMS–CMSC–NAIMS consensus recommendations on the use of MRI for the diagnosis of multiple sclerosis, Barkhof et al., Lancet Neurol 2025; 24: 866–79
The 3D T2* gradient echo that we use on Philips Achieva in Nottingham & on Siemens Magnetom Prisma in Cardiff has the following specifications:
sagittal acquisition, 0.6 × 0.6 × 0.6 mm
voxel size, 230 x 230 x 180 mm field of view, effective echo
time 25 ms
repetition time of 55 ms
parallel imaging factor 2
10-degree flip angle
echo planar imaging factor or multiecho options if available
scan duration of 6 minutes or less
Here is further information for Philips users: